Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
Published June 26, 2013 · Recall initiated: May 15, 2013
Summary
FDA records medical device Class II recall Z-1548-2013, published June 26, 2013 and initiated May 15, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 17; Lot: Catalog number: 010001814 and lot numbers: 318520 and 644720. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.
- Lot
- Catalog number: 010001814 and lot numbers: 318520 and 644720.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDABiomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The PhoenixJune 26, 2013
- FDAAperio ScanScope System. Aperio ScanScope Models, XT, AT, AT Turbo, CS and CSO. Automated digital slideJune 26, 2013
Questions about this recall
Was Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain. recalled?
Yes. FDA published recall Z-1548-2013 on June 26, 2013. Reported hazard: Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.
What should I do if I have this product?
Voluntary: Firm initiated · Letter