Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
Published June 26, 2013 · Recall initiated: May 15, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.
- Lot
- Catalog number: 010001814 and lot numbers: 318520 and 644720.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States