Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM

Published May 17, 2023 · Recall initiated: April 11, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue, which could lead to procedure delay, and insufficient sample acquisition requiring a repeat procedure.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Bard Peripheral Vascular Inc
Distributor
US Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY . OUS: BE, CA, AU
Lot
REF/UDI-DI/Lot (Expiration Date): 2010MSK/00801741097089/0001445946(11/28/2024), 2016MSK/00801741097096/1458844(02/28/2025)
Risk classification
Class II