Refill Kit 8551/856X (includes Models 8561, 8562 and 8564) for use with Medtronic Implantable Programmable Infusion Pumps. Product Usage: The Refill Kits are intended for use in refilling Medtronic programmable, implantable infusion pumps with the exception of Medtronic MiniMed infusion pumps.
Published May 14, 2014 · Recall initiated: April 4, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some refill kits in the affected lots may contain an extension tubing set that is occluded and healthcare professionals may be unable to aspirate or fill the SynchroMed pumps. There is no problem with the Lioresal Intrathecal (baclofen injection) contained in the affected drug refill kits. There have been no patient injuries beyond additional needle sticks associated with this issue.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Medtronic Neuromodulation
- Distributor
- Worldwide Distribution - USA Nationwide in the states of NM, ND, RI, and VT and the countries of Australia, Singapore, Thailand, Central and Eastern Europe, Western Europe, Middle East, & Panama.
- Lot
- Lot numbers CS3643, CS3613, CS3663, N441998, N442081, N442131, N440259, N440527, N440582, N440654, N440909, N440978, N441199, N441268, N441806, N441817, N442178, N442513, N442532, N442560, N442588, N442831, and N443427. Affected lots were distributed between January 2014 and March 2014.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States