Medtronic SynchroMed¿ II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.
Published May 14, 2014 · Recall initiated: February 26, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Medtronic Neuromodulation
- Distributor
- Worldwide Distribution - All states in USA. OUS: List not provided at this time.
- Lot
- This Medical Device Correction notification affects all SynchroMed II pumps.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States