INFUSE(R) Bone Graft,LARGE II KIT, REF 7510800, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
Published July 3, 2013 · Recall initiated: May 28, 2013
Summary
FDA records medical device Class II recall Z-1592-2013, published July 3, 2013 and initiated May 28, 2013. Recalling firm: Medtronic Sofamor Danek USA Inc; Status: Terminated; Reported quantity: 16994 units; Lot: Lots: M111103AAG, M111052AAQ, M111052AAR, M111052AAP, M111052AAL, M111059AAC, M111059AAF, M111052AAF, M111059AAH, M111103AAB, M111059AAG, M111052AAK, M111052AAJ, M111059AAB, M111059AAK, M111052AAI, M111052AAO, M111052AAH, M111052AAG, M111059AAD. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Sofamor Danek USA Inc
- Distributor
- Worldwide distribution: USA (nationwide including Puerto Rico) and Internationally to Australia, Brazil, Canada, Chile, Hong Kong, India, Mexico, Netherlands, , New South Wales, Panama, and Paraguay.
- Lot
- Lots: M111103AAG, M111052AAQ, M111052AAR, M111052AAP, M111052AAL, M111059AAC, M111059AAF, M111052AAF, M111059AAH, M111103AAB, M111059AAG, M111052AAK, M111052AAJ, M111059AAB, M111059AAK, M111052AAI, M111052AAO, M111052AAH, M111052AAG, M111059AAD
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGrafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive toOctober 21, 2026
- FDAGrafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive toOctober 21, 2026
- FDAGrafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone MatrixNovember 1, 2023
- FDAEndoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody,October 20, 2021
- FDAINFUSE(R) Bone Graft, LARGE KIT, REF 7510600, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis,July 3, 2013
- FDAINFUSE(R) Bone Graft, SMALL KIT, REF 7510200AUS, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place,July 3, 2013
Questions about this recall
Was INFUSE(R) Bone Graft,LARGE II KIT, REF 7510800, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic recalled?
Yes. FDA published recall Z-1592-2013 on July 3, 2013. Reported hazard: Kits contain a collagen sponge recalled by Integra due to possible elevated endotoxin levels resulting from a process deviation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter