FDAFDA · Food and drugs·Recall number Z-1604-2018

708033 URODiagnost As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Published May 9, 2018 · Recall initiated: March 26, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Medical Systems Nederlands
Distributor
Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Lot
System Serial # 8 14 1 64 15 4 3 7 18 48 16 2 59 60 17 11 13 21 5
Risk classification
Class II
Area (FDA)
device
Country
Netherlands