Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.

Published April 8, 2020 · Recall initiated: February 20, 2020

Summary

FDA records medical device Class II recall Z-1616-2020, published April 8, 2020 and initiated February 20, 2020. Recalling firm: Philips North America, LLC; Status: Terminated; Reported quantity: 2214 units; Lot: Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America, LLC
Distributor
Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.
Lot
Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner. recalled?

Yes. FDA published recall Z-1616-2020 on April 8, 2020. Reported hazard: Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00

What should I do if I have this product?

Voluntary: Firm initiated