Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.
Published April 8, 2020 · Recall initiated: February 20, 2020
Summary
FDA records medical device Class II recall Z-1616-2020, published April 8, 2020 and initiated February 20, 2020. Recalling firm: Philips North America, LLC; Status: Terminated; Reported quantity: 2214 units; Lot: Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America, LLC
- Distributor
- Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.
- Lot
- Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for theApril 20, 2022
- FDAPhilips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A,November 4, 2020
- FDAPhilips SureSigns VM1SpO2, C02, Rec, model no. 863266 - Product Usage: The SureSigns VM1 is a patient monitorJune 17, 2020
- FDAPhilips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospitalMay 27, 2020
- FDAInstructions for Use and Surgical Cleaning and Sterilization Protocol for Trilliant Surgical Ltd GridlockApril 8, 2020
- FDAKarl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series,April 8, 2020
Questions about this recall
Was Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner. recalled?
Yes. FDA published recall Z-1616-2020 on April 8, 2020. Reported hazard: Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00
What should I do if I have this product?
Voluntary: Firm initiated