Homocysteine test system - Homocysteine Reagent. 05385415190; Homocysteine 100 Tests, cobas c, COBAS INTEGRA Homocysteine Enzymatic Assay is an in vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems.
Published May 27, 2015 · Recall initiated: March 23, 2015
Summary
FDA records medical device Class II recall Z-1622-2015, published May 27, 2015 and initiated March 23, 2015. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 2558 units; Lot: 05385415190, Homocysteine 100Tests, cobas c, COBAS INTEGRA Lot number 69781101. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customers complained about under-recovery of non-Roche controls and discrepant low patient results with certain Homocysteine reagent lots. This negative bias could, in the worst case, lead to inaccurately low Homocysteine results. An elevated level of Homocysteine is considered an important risk factor in the assessment of peripheral vascular disease. Falsely low values could lead to a delay of di
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- Nationwide Distribution-including PR and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, , SC, TN, TX, VA, WA, and WV.
- Lot
- 05385415190, Homocysteine 100Tests, cobas c, COBAS INTEGRA Lot number 69781101
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, inNovember 27, 2024
- FDACreatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitativeOctober 2, 2024
- FDAAlbumin Gen.2, Catalog 05166861190October 21, 2020
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virusDecember 18, 2019
- FDAMedtronic DBS Extension Kit for Deep Brain Stimulation, Models 37085 and 37086. SterileMay 27, 2015
- FDAPHALINX (TM) HAMMERTOE HANDLE, SIZE: LARGE, STYLE: ANGLED, REF 45304014, CONTENTS 1 EACH, RX ONLYMay 27, 2015
Questions about this recall
Was Homocysteine test system - Homocysteine Reagent. 05385415190; Homocysteine 100 Tests, cobas c, COBAS INTEGRA Homocysteine Enzymatic Assay is an in vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems. recalled?
Yes. FDA published recall Z-1622-2015 on May 27, 2015. Reported hazard: Customers complained about under-recovery of non-Roche controls and discrepant low patient results with certain Homocysteine reagent lots. This negative bias could, in the worst case, lead to inaccurately low Homocysteine results. An elevated level of Homocysteine is considered an important risk factor in the assessment of peripheral vascular disease. Falsely low values could lead to a delay of di
What should I do if I have this product?
Voluntary: Firm initiated · Letter