Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.
Published May 28, 2014 · Recall initiated: April 4, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Medical Systems, Inc.
- Distributor
- Worldwide Distribution - Including US Nationwide, Canada, Czech Republic, Denmark, Netherlands, Estonia, Finland, France, Germany, Greece, Hungary, Italy, Japan, Kazakhstan, Norway, Poland, Romania, Russia, China, Slovakia, Sweden, and Turkey.
- Lot
- Patient Monitor Product numbers M3002A, M8102A, M8105A, M8001A, M8002A, M8003A, M8004A, M8027A, M8001AU, M8002AU, M8003AU, and M8004AU with software releases up to and including G.0.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States