Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.
Published May 28, 2014 · Recall initiated: April 4, 2014
Summary
FDA records medical device Class II recall Z-1628-2014, published May 28, 2014 and initiated April 4, 2014. Recalling firm: Philips Medical Systems, Inc.; Status: Terminated; Reported quantity: 188,520 devices; Lot: Patient Monitor Product numbers M3002A, M8102A, M8105A, M8001A, M8002A, M8003A, M8004A, M8027A, M8001AU, M8002AU, M8003AU, and M8004AU with software releases up to and including G.0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems, Inc.
- Distributor
- Worldwide Distribution - Including US Nationwide, Canada, Czech Republic, Denmark, Netherlands, Estonia, Finland, France, Germany, Greece, Hungary, Italy, Japan, Kazakhstan, Norway, Poland, Romania, Russia, China, Slovakia, Sweden, and Turkey.
- Lot
- Patient Monitor Product numbers M3002A, M8102A, M8105A, M8001A, M8002A, M8003A, M8004A, M8027A, M8001AU, M8002AU, M8003AU, and M8004AU with software releases up to and including G.0.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFoot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 ThisApril 6, 2016
- FDAPhilips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDALATEX-FREE; 4-LEAD T-U-R IRRIGATION SET; Nonvented Set for Transurethral Resection Procedures; 85 inch (216May 28, 2014
- FDAAnspach Irrigation Clip 40 Product Usage: Irrigation clips provide a method of supplying controlled, coolingMay 28, 2014
Questions about this recall
Was Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters. recalled?
Yes. FDA published recall Z-1628-2014 on May 28, 2014. Reported hazard: The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.
What should I do if I have this product?
Voluntary: Firm initiated · Letter