Arrowgard Blue and Arrowgard Blue Plus Pressure Injectable Central Venous Catheter, a multiple-lumen catheter that permits venous access to central circulation. Catalog number ASK-45703-PHF.
Published May 28, 2014 · Recall initiated: July 17, 2013
Summary
FDA records medical device Class II recall Z-1633-2014, published May 28, 2014 and initiated July 17, 2013. Recalling firm: Arrow International Inc; Status: Terminated; Reported quantity: 4285; Lot: MF2016829 RM2072044 RM2034076 RM2095792 RM2056629 RM3015571 RM2069733. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The labeling states that the kit contains Introducer Needle: Echogenic 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe; however, the kit contains, and should state, Introducer Needle: 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe. As a result of this labeling inconsistency, if the clinician were to place the needle under x-r
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Arrow International Inc
- Distributor
- US Distribution only to MI.
- Lot
- MF2016829 RM2072044 RM2034076 RM2095792 RM2056629 RM3015571 RM2069733
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500JJuly 8, 2020
- FDAArrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701July 8, 2020
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500July 8, 2020
- FDAArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400July 8, 2020
- FDAMaXair True Low Air Loss Mattress Replacement System, True Air Technologies, Inc.May 28, 2014
- FDACDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers -May 28, 2014
Questions about this recall
Was Arrowgard Blue and Arrowgard Blue Plus Pressure Injectable Central Venous Catheter, a multiple-lumen catheter that permits venous access to central circulation. Catalog number ASK-45703-PHF. recalled?
Yes. FDA published recall Z-1633-2014 on May 28, 2014. Reported hazard: The labeling states that the kit contains Introducer Needle: Echogenic 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe; however, the kit contains, and should state, Introducer Needle: 18 Ga. x 2.5 (6.35cm) XTW and 5mL ArroW Raulerson Spring-Wire Introduction Syringe. As a result of this labeling inconsistency, if the clinician were to place the needle under x-r
What should I do if I have this product?
Voluntary: Firm initiated · Letter