ORTHO® Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO® Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing.
Published July 10, 2013 · Recall initiated: May 6, 2013
Summary
FDA records medical device Class II recall Z-1651-2013, published July 10, 2013 and initiated May 6, 2013. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: 31 units; Lot: Serial numbers: 0811F06216 through 0811F06225 and 0911F06240 through 0911F06245 and 1011F06317 through 1011F06331. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
- Lot
- Serial numbers: 0811F06216 through 0811F06225 and 0911F06240 through 0911F06245 and 1011F06317 through 1011F06331
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDAVITROS 3600 Immunodiagnostic System (product code 6802783) Product Usage: For use in the in vitroJuly 10, 2013
- FDALithium Thionyl non-rechargeable disposable battery pack P/N 350-0294-01 used with ER920 Wireless SeriesJuly 10, 2013
Questions about this recall
Was ORTHO® Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO® Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing. recalled?
Yes. FDA published recall Z-1651-2013 on July 10, 2013. Reported hazard: There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
What should I do if I have this product?
Voluntary: Firm initiated · Letter