***REF SC2102***eMax 2 Plus Console, Single Port, no Irrigation***SN G16307882402*** Product Usage: The eMax console is intended to operate the eMax system which is used for cutting and shaping bone including spine and cranium. The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM).

Published July 10, 2013 · Recall initiated: June 24, 2011

Summary

FDA records medical device Class II recall Z-1659-2013, published July 10, 2013 and initiated June 24, 2011. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 61 units; Lot: SC2100 System Console Emax 2 Plus; SC2101 System Console, Single Port with irrigation; SC2102 System Console, Single Port, without Irrigation. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The RPM of the Motor Handpiece may be less than 80,000 RPM.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
The Anspach Effort, Inc.
Distributor
Worldwide Distribution - USA Nationwide including the states of: AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of: Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;
Lot
SC2100 System Console Emax 2 Plus; SC2101 System Console, Single Port with irrigation; SC2102 System Console, Single Port, without Irrigation;
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ***REF SC2102***eMax 2 Plus Console, Single Port, no Irrigation***SN G16307882402*** Product Usage: The eMax console is intended to operate the eMax system which is used for cutting and shaping bone including spine and cranium. The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM). recalled?

Yes. FDA published recall Z-1659-2013 on July 10, 2013. Reported hazard: The RPM of the Motor Handpiece may be less than 80,000 RPM.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit