FDAFDA · Food and drugs·Recall number Z-1684-2022
Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F)
Published September 14, 2022 · Recall initiated: August 10, 2022
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check before use ensure that the device is functioning appropriately.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Saranas, Inc.
- Distributor
- US Nationwide distribution including in the states of PA, MI, FL, WV, TX, NJ, GA, LA, NY, MA, CT, AZ, OH, TN, VA.
- Lot
- UDI/REF/Lot: 00866532000402/FG001-01/510090S, 511110A, 510120F, 102422, 102445, 102531, 102671, 102685, 102731; 00866532000419/FG001-02/515070A, 528090C, 510120G, 511110B, 102567, 102610, 102611
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States