OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0
Published May 8, 2024 · Recall initiated: March 11, 2024
Summary
FDA records medical device Class II recall Z-1696-2024, published May 8, 2024 and initiated March 11, 2024. Recalling firm: Advanced Instruments, LLC; Status: Ongoing; Reported quantity: 58 units; Lot: UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A. All serial numbers that end in A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Advanced Instruments, LLC
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, France, Germany, Iceland, Netherlands, South Korea, Spain, Switzerland, Taiwan.
- Lot
- UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A. All serial numbers that end in A
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAbbott Aveir Leadless Pacemaker, REF: LSP112V, STERILEEOMay 8, 2024
- FDASofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical sutureMay 8, 2024
- FDAKeller Funnel 2, REF: HA-001, STERILEEO, Rx Only,May 8, 2024
- FDAADVANCED RESEARCH MEDICAL Lumbar Interbody Fusion System (OLLIF), Flexible Curette Loop Blade, Part # 01-06-4;May 8, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnlyMay 8, 2024
- FDAPOWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgeryMay 8, 2024
Questions about this recall
Was OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0 recalled?
Yes. FDA published recall Z-1696-2024 on May 8, 2024. Reported hazard: System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail