POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0537CJDA
Published May 8, 2024 · Recall initiated: March 7, 2024
Summary
FDA records medical device Class II recall Z-1700-2024, published May 8, 2024 and initiated March 7, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 850 units: (400 units US);(450 units OUS); Lot: UDI-DI: 00821925044555 Lot numbers: CA182543, CA182544, CA182546, CA182548, CA191230. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.
- Lot
- UDI-DI: 00821925044555 Lot numbers: CA182543, CA182544, CA182546, CA182548, CA191230
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDAPOWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgeryMay 8, 2024
- FDAPOWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgeryMay 8, 2024
Questions about this recall
Was POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0537CJDA recalled?
Yes. FDA published recall Z-1700-2024 on May 8, 2024. Reported hazard: Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit