The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.
Published April 22, 2020 · Recall initiated: March 9, 2020
Summary
FDA records medical device Class II recall Z-1703-2020, published April 22, 2020 and initiated March 9, 2020. Recalling firm: Sysmex America, Inc.; Status: Terminated; Reported quantity: 93 units.; Lot: Catalog No. AH699752. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sysmex America, Inc.
- Distributor
- USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..
- Lot
- Catalog No. AH699752.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10April 22, 2020
- FDASPICE (JWH-018) CC443April 22, 2020
- FDAVACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.April 22, 2020
- FDAThe UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. TheFebruary 19, 2014
- FDASysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,November 14, 2012
- FDASysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,November 14, 2012
Questions about this recall
Was The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines. recalled?
Yes. FDA published recall Z-1703-2020 on April 22, 2020. Reported hazard: Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.
What should I do if I have this product?
Voluntary: Firm initiated · Letter