The CG+ Arrow PIC Catheters permit venous access to the central circulation through a peripheral vein.
Published April 12, 2017 · Recall initiated: March 3, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Arrow International, Inc. notified its customers who received the affected Antimicrobial/Antithrombogenic PICC Kits, that the kits were being recalled due to a labeling error. The lidstock on some of the kits reads CDC-45541-HPK1A, while the banner card on the side of the kit reads CDC-45052-HPK1A. Other kits may list CDC-45052-HPK1A on the lidstock and CDC-45541-HPK1A on the banner card. The affected lot numbers are 23F16C0851, 23F16A0727, and 23F16C0844.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Arrow International Inc
- Distributor
- US Distribution
- Lot
- Device Listing Number: D149129 Lot Numbers: 23F16C0851, 23F16A0727 and 23F16C0844
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?