OMNIBotics iBlock Cutting Guide, 4146-1000 - Product Usage: The OMNIBotics iBlock Cutting Guide is a stainless-steel surgical instrument used during total knee arthroplasty. Specifically, the iBlock Cutting Guide is used in conjunction with the iBlock Saw Guide with the OMNIBot robotic Motor Unit, and for preparation of the bone for implant implantation. The instrument has been designed to facilitate bone preparation and implantation of the OMNI Apex Knee System.

Published June 12, 2019 · Recall initiated: April 12, 2019

Summary

FDA records medical device Class II recall Z-1717-2019, published June 12, 2019 and initiated April 12, 2019. Recalling firm: OMNIlife science Inc.; Status: Terminated; Reported quantity: 9; Lot: MI1218. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during use.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
OMNIlife science Inc.
Distributor
US Nationwide distribution in the states of OK, SC, and WI.
Lot
MI1218
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OMNIBotics iBlock Cutting Guide, 4146-1000 - Product Usage: The OMNIBotics iBlock Cutting Guide is a stainless-steel surgical instrument used during total knee arthroplasty. Specifically, the iBlock Cutting Guide is used in conjunction with the iBlock Saw Guide with the OMNIBot robotic Motor Unit, and for preparation of the bone for implant implantation. The instrument has been designed to facilitate bone preparation and implantation of the OMNI Apex Knee System. recalled?

Yes. FDA published recall Z-1717-2019 on June 12, 2019. Reported hazard: The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during use.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit