Synthes CMF Distraction System The product is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus, where gradual bone distraction is required. It is intended for single use only.
Published July 24, 2013 · Recall initiated: June 12, 2013
Summary
FDA records medical device Class II recall Z-1726-2013, published July 24, 2013 and initiated June 12, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 10; Lot: Part # 04.315.067 with lot #s 6184270 and/or 22612-04. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF) Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots in which the slot in the barrel of the BC distractor was rotated such that it was not possible to attach the B-type and C-type foot plates.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Nationwide Distribution including PA, CT, DE, and NY.
- Lot
- Part # 04.315.067 with lot #s 6184270 and/or 22612-04.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedgeDecember 18, 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDAMedStream Programmable Infusion Pump , Implantable for the intrathecal delivery of Baclofen or MorphineJuly 24, 2013
- FDAB-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitativeJuly 24, 2013
Questions about this recall
Was Synthes CMF Distraction System The product is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus, where gradual bone distraction is required. It is intended for single use only. recalled?
Yes. FDA published recall Z-1726-2013 on July 24, 2013. Reported hazard: The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF) Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots in which the slot in the barrel of the BC distractor was rotated such that it was not possible to attach the B-type and C-type foot plates.
What should I do if I have this product?
Voluntary: Firm initiated · Letter