FDAFDA · Food and drugs·Recall number Z-1726-2013
Synthes CMF Distraction System The product is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus, where gradual bone distraction is required. It is intended for single use only.
Published July 24, 2013 · Recall initiated: June 12, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF) Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots in which the slot in the barrel of the BC distractor was rotated such that it was not possible to attach the B-type and C-type foot plates.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Nationwide Distribution including PA, CT, DE, and NY.
- Lot
- Part # 04.315.067 with lot #s 6184270 and/or 22612-04.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?