Back to search
FDAFDA · Food and drugs·Recall number Z-1726-2013

Synthes CMF Distraction System The product is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus, where gradual bone distraction is required. It is intended for single use only.

Published July 24, 2013 · Recall initiated: June 12, 2013

Summary

FDA records medical device Class II recall Z-1726-2013, published July 24, 2013 and initiated June 12, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 10; Lot: Part # 04.315.067 with lot #s 6184270 and/or 22612-04. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF) Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots in which the slot in the barrel of the BC distractor was rotated such that it was not possible to attach the B-type and C-type foot plates.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Synthes USA HQ, Inc.
Distributor
Nationwide Distribution including PA, CT, DE, and NY.
Lot
Part # 04.315.067 with lot #s 6184270 and/or 22612-04.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Synthes CMF Distraction System The product is intended for use as a bone stabilizer and lengthening (and/or transport) device for correction of congenital deficiencies or post-traumatic defects of the mandibular body and ramus, where gradual bone distraction is required. It is intended for single use only. recalled?

Yes. FDA published recall Z-1726-2013 on July 24, 2013. Reported hazard: The firm initiated a voluntary recall of the BC Distractor Body which is part of the Craniomaxillofacial (CMF) Distractor System, due to a mis-alignment issue. There have been instances reported within the impacted lots in which the slot in the barrel of the BC distractor was rotated such that it was not possible to attach the B-type and C-type foot plates.

What should I do if I have this product?

Voluntary: Firm initiated · Letter