FDAFDA · Food and drugs·Recall number Z-1727-2013
B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals diagnosed with osteopenia. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.
Published July 24, 2013 · Recall initiated: April 25, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Roche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys ¿-CrossLaps (¿-CL) assay package insert version 5, could not be reproduced. The ¿-CrossLaps test results contribute to the monitoring of antiresorptive therapy. False low or false normal results may lead to delayed diagnosis and treatment of osteoporosi
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- Nationwide distribution: USA including states of: AL, CA, GA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, TX, VA, VT, and WA.
- Lot
- All lot numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States