Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia.
Published July 24, 2013 · Recall initiated: May 9, 2013
Summary
FDA records medical device Class II recall Z-1736-2013, published July 24, 2013 and initiated May 9, 2013. Recalling firm: Integra LifeSciences Corporation; Status: Terminated; Reported quantity: 692; Lot: All lot numbers for PANTA Arthrodesis Nail Support Device Catalogue Nos: 519-110 and 519-130 from May 2006 to current. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during use of the Panta Nail Support Device, Integra has decided to issue this voluntary medical device correction. The misalignment can result in the inability to properly insert the screws into the Panta Nail and create the potential need for revision surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Integra LifeSciences Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Israel, Italy, Netherlands, Portugal, Russia, Saudia Arabia, South Africa, Spain and United Arab Emirates.
- Lot
- All lot numbers for PANTA Arthrodesis Nail Support Device Catalogue Nos: 519-110 and 519-130 from May 2006 to current.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntegra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra® DermalSeptember 21, 2016
- FDAIntegra Newdeal SURFIX Fixed Angle Locking Screw 3.5mm x 22mm (catalog number 285322S) and 3.5mm x 30mmJune 11, 2014
- FDAVarious types of Fetal Bovine Serum Bovine sourced Animal Sera for in-vitro diagnostic device and as a growthJuly 24, 2013
- FDAKRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as aJuly 24, 2013
- FDAOra Plug Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings forMay 22, 2013
- FDAHeliTape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings forMay 22, 2013
Questions about this recall
Was Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. recalled?
Yes. FDA published recall Z-1736-2013 on July 24, 2013. Reported hazard: As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during use of the Panta Nail Support Device, Integra has decided to issue this voluntary medical device correction. The misalignment can result in the inability to properly insert the screws into the Panta Nail and create the potential need for revision surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter