FDAFDA · Food and drugs·Recall number Z-1740-2024

The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. Device dimensions and configuration are shown on the product label. A rotating hemostasis valve with side-arm adapter is provided with each catheter. The rotating hemostasis valve is typically y shaped with a female luer lock and a manual hemostasis valve. The female port allows for aspiration and contrast injections while the hemostasis valve allows direct arterial access when using other devices such as guidewires and/or interventional devices.

Published May 8, 2024 · Recall initiated: April 3, 2024

Summary

FDA records medical device Class II recall Z-1740-2024, published May 8, 2024 and initiated April 3, 2024. Recalling firm: Stryker Neurovascular; Status: Ongoing; Reported quantity: 43; Lot: UDI-DI: 07613327313895, Lot: 0000486382, Expiration: 17-Jul-2025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Neurovascular
Distributor
US Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA, MI, MS, SC, IL.
Lot
UDI-DI: 07613327313895, Lot: 0000486382, Expiration: 17-Jul-2025
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The Distal Access Catheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. Device dimensions and configuration are shown on the product label. A rotating hemostasis valve with side-arm adapter is provided with each catheter. The rotating hemostasis valve is typically y shaped with a female luer lock and a manual hemostasis valve. The female port allows for aspiration and contrast injections while the hemostasis valve allows direct arterial access when using other devices such as guidewires and/or interventional devices. recalled?

Yes. FDA published recall Z-1740-2024 on May 8, 2024. Reported hazard: Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter