Back to search
FDAFDA · Food and drugs·Recall number Z-1751-2013

LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the administration of I.V. fluids; Made in Costa Rica; Hospira, Inc., Lake Forest, IL 60045 USA The Hospira Symbiq Infusion System is a general infusion system which provides IV infusion across the general spectrum of clinical care. The Symbiq pumps are standard sized pumps which are typically facility-based infusion pumps which may be used to deliver continuous and/or intermittent infusions to a variety of patient populations.

Published July 24, 2013 · Recall initiated: June 18, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve. Preliminary investigation findings found that there is a potential for the ball valve to not seal the drip chamber as intended, leading to air in the line of the administration set.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hospira Inc.
Distributor
Worldwide Distribution - US (nationwide) including the states of IA, IL, MN, SD and WA., and Costa Rica.
Lot
List Number: 19666-04-28; Lot Number: 87-188-5H
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card