FDAFDA · Food and drugs·Recall number Z-1760-2020

ABL90 FLEX Analyzer, Model Numbers 393-090, with software versions below 3.4MR2 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.

Published April 29, 2020 · Recall initiated: February 3, 2020

Summary

FDA records medical device Class II recall Z-1760-2020, published April 29, 2020 and initiated February 3, 2020. Recalling firm: Radiometer Medical ApS; Status: Terminated; Reported quantity: 12,822 units; Lot: software versions below 3.4 MR2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Radiometer Medical ApS
Distributor
Worldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada, Netherlands, China, Croatia, Czech Republic, Denmark, Germany, Spain, France, Hungary, India, Japan, South Korea, United Kingdom, Poland, Australia, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland.
Lot
software versions below 3.4 MR2
Risk classification
Class II
Area (FDA)
device
Country
Denmark

Related recalls

Questions about this recall

Was ABL90 FLEX Analyzer, Model Numbers 393-090, with software versions below 3.4MR2 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting. recalled?

Yes. FDA published recall Z-1760-2020 on April 29, 2020. Reported hazard: The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.

What should I do if I have this product?

Voluntary: Firm initiated · Letter