Back to search
FDAFDA · Food and drugs·Recall number Z-1766-2017

The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis of the radial and ulnar artery following catheterization procedures. It has a soft wristband with a secure hook and loop fastener. The band delivers adjustable compression of the puncture site with an inflatable bulb, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement.

Published April 19, 2017 · Recall initiated: March 3, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may enter the valve, preventing a complete seal of the inflation valve.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Merit Medical Systems, Inc.
Distributor
Nationwide, Canadian and other foreign consignees. No VA/govt/military consignees.
Lot
Catalog Number: SRB24AC, Lot Numbers: H1068590S1, H1078177, H1078180, H1083628; Catalog Number: SRB24MED, Lot Numbers: H1068587S1,H1073974, H1078179, H1078181, H1083629, H1083630; Catalog Number: SRB29AC, Lot Numbers: H1068591S1, H1078183, H1083631; Catalog Number: SRB29MED, Lot Numbers: H1068589S1, H1078184, H1078185, H1083633.
Risk classification
Class II
Area (FDA)
device
Country
United States