PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.
Published July 31, 2013 · Recall initiated: March 27, 2012
Summary
FDA records medical device Class II recall Z-1774-2013, published July 31, 2013 and initiated March 27, 2012. Recalling firm: Topcon Medical Laser Systems, Inc; Status: Terminated; Reported quantity: 7 US, 33 foreign; Lot: PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x Serial numbers: United States 33057910 33060410 33063610 66078611 66084911 66085111 66085511 Out of United States OMC-179 11138509 11141209 11147209 11147809 11148910 11149010 11149210 11149710 . 11149810 11150110 11150210 11150410 11150610 11150810 11150910 11151010 1 151110 11151211 33054310 33059310 33059510 33062810" 66075211' 66076711 66077111 66078011 66079911 6608391'1 66084611 66084811 66085011 66085211. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Topcon Medical Laser Systems, Inc
- Distributor
- Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.
- Lot
- PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x Serial numbers: United States 33057910 33060410 33063610 66078611 66084911 66085111 66085511 Out of United States OMC-179 11138509 11141209 11147209 11147809 11148910 11149010 11149210 11149710 . 11149810 11150110 11150210 11150410 11150610 11150810 11150910 11151010 1 151110 11151211 33054310 33059310 33059510 33062810" 66075211' 66076711 66077111 66078011 66079911 6608391'1 66084611 66084811 66085011 66085211
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe LIFEPAK 500 Automated External Defibrillator is a semi-automatic defibrillator that uses a patented ShockJuly 31, 2013
- FDALIGACLIP® Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Product Usage: The LIGACLIP EndoscopicJuly 31, 2013
- FDAEndopath Electrosurgery Probe Plus II, twelve (12) different devices, including hooks, spatulas, right angles,July 31, 2013
- FDALIGACLIP® Endoscopic Rotating Multiple Clip Applier Convenience Kit (Flex Trays), 10mm M/L Product Usage: TheJuly 31, 2013
- FDASpacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery ofJuly 31, 2013
- FDALIGACLIP® Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L, bulk non-sterile Product Usage: TheJuly 31, 2013
Questions about this recall
Was PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye. recalled?
Yes. FDA published recall Z-1774-2013 on July 31, 2013. Reported hazard: When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter