Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
Published May 15, 2024 · Recall initiated: April 17, 2024
Summary
FDA records medical device Class II recall Z-1784-2024, published May 15, 2024 and initiated April 17, 2024. Recalling firm: NRT X-RAY A/S; Status: Ongoing; Reported quantity: 35; Lot: REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- NRT X-RAY A/S
- Distributor
- US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
- Lot
- REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Denmark
Related recalls
- FDAGE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAStryker 1,5/1,7MM DYNAMIC MESH- STD-MEDIUM-Intended for the reconstruction, stabilization and rigid fixationMay 15, 2024
- FDAGE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drugMay 15, 2024
- FDAGE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systemsMay 15, 2024
- FDAIntelli-C, Left side suspended Tabletop, REF: 03400010June 7, 2023
Questions about this recall
Was Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system recalled?
Yes. FDA published recall Z-1784-2024 on May 15, 2024. Reported hazard: X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
What should I do if I have this product?
Voluntary: Firm initiated · Letter