STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.

Published July 31, 2013 · Recall initiated: May 20, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati

What to do

Voluntary: Firm initiated · Visit

Recall details

Manufacturer
Steris Corporation
Distributor
USA Nationwide Distribution in the states of CO, KY, LA, VA, and WI. The affected product was also distributed to Canada.
Lot
The following serial/lot numbers are affected by this recall: 0404913150, 0404913151, 0404913152, 0404913153, 0408713001, 0408713002, 0408713003, 0408713004, 0408713005, 0408713006, 0409813097, 0409813098, 0409813099, 0409813100, 0410113001, 0410113002, 0410713024, 0410713025, 0410713026, 0410713027, 0410713028, and 0410713029.
Risk classification
Class II
Area (FDA)
device
Country
United States