STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.
Published July 31, 2013 · Recall initiated: May 20, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati
What to do
Voluntary: Firm initiated · Visit
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- USA Nationwide Distribution in the states of CO, KY, LA, VA, and WI. The affected product was also distributed to Canada.
- Lot
- The following serial/lot numbers are affected by this recall: 0404913150, 0404913151, 0404913152, 0404913153, 0408713001, 0408713002, 0408713003, 0408713004, 0408713005, 0408713006, 0409813097, 0409813098, 0409813099, 0409813100, 0410113001, 0410113002, 0410713024, 0410713025, 0410713026, 0410713027, 0410713028, and 0410713029.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States