FDAFDA · Food and drugs·Recall number Z-1787-2012
GE Healthcare, Aestiva/5 7900 SmartVent, anesthesia machine. Used in Datex-Ohmeda Aestiva/5 Anesthesia Systems.
Published June 27, 2012 · Recall initiated: April 10, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Nationwide Distribution - including the states of AL, FL, GA, MO, and VA.
- Lot
- Serial Number AMRP01031 AMRP00966 AMRP01030 AMRP00968 AMRP00967 AMRP01033 AMRP00970 AMRP00969
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States