Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Published June 25, 2014 · Recall initiated: April 29, 2014

Summary

FDA records medical device Class II recall Z-1789-2014, published June 25, 2014 and initiated April 29, 2014. Recalling firm: Childrens Medical Ventures; Status: Terminated; Reported quantity: 112; Lot: 989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Childrens Medical Ventures
Distributor
Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.
Lot
989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor. recalled?

Yes. FDA published recall Z-1789-2014 on June 25, 2014. Reported hazard: The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.