FDAFDA · Food and drugs·Recall number Z-1789-2014
Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
Published June 25, 2014 · Recall initiated: April 29, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Childrens Medical Ventures
- Distributor
- Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.
- Lot
- 989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States