FDAFDA · Food and drugs·Recall number Z-1793-2012

Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplastics or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture

Published June 20, 2012 · Recall initiated: February 21, 2012

Summary

FDA records medical device Class II recall Z-1793-2012, published June 20, 2012 and initiated February 21, 2012. Recalling firm: Aesculap Implant Systems; Status: Terminated; Reported quantity: 8; Lot: NR068K, TIBIA PLATEAU, FIXED, T0; NR070K, TIBIA PLATEAU, FIXED, T0+; NR072K, TIBIA PLATEAU, FIXED, T1; NR072K, TIBIA PLATEAU, FIXED, T1+; NR073K, TIBIA PLATEAU, FIXED, T2; NR074K, TIBIA PLATEAU, FIXED, T2+; NR075K, TIBIA PLATEAU, FIXED, T3; NR076K, TIBIA PLATEAU, FIXED, T3+; NR077K, TIBIA PLATEAU, FIXED, T4; NR078K, TIBIA PLATEAU, FIXED, T4+; NR079K, TIBIA PLATEAU, FIXED, T5. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aesculap Implant Systems
Distributor
Nationwide Distribution including PA and TX.
Lot
NR068K, TIBIA PLATEAU, FIXED, T0; NR070K, TIBIA PLATEAU, FIXED, T0+; NR072K, TIBIA PLATEAU, FIXED, T1; NR072K, TIBIA PLATEAU, FIXED, T1+; NR073K, TIBIA PLATEAU, FIXED, T2; NR074K, TIBIA PLATEAU, FIXED, T2+; NR075K, TIBIA PLATEAU, FIXED, T3; NR076K, TIBIA PLATEAU, FIXED, T3+; NR077K, TIBIA PLATEAU, FIXED, T4; NR078K, TIBIA PLATEAU, FIXED, T4+; NR079K, TIBIA PLATEAU, FIXED, T5
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplastics or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture recalled?

Yes. FDA published recall Z-1793-2012 on June 20, 2012. Reported hazard: During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem.

What should I do if I have this product?

Voluntary: Firm initiated · Letter