Back to search
FDAFDA · Food and drugs·Recall number Z-1796-2020

PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

Published May 6, 2020 · Recall initiated: March 11, 2020

Summary

FDA records medical device Class II recall Z-1796-2020, published May 6, 2020 and initiated March 11, 2020. Recalling firm: Merit Medical Systems, Inc.; Status: Terminated; Reported quantity: 1932; Lot: SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Merit Medical Systems, Inc.
Distributor
U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
Lot
SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT recalled?

Yes. FDA published recall Z-1796-2020 on May 6, 2020. Reported hazard: During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.

What should I do if I have this product?

Voluntary: Firm initiated · Letter