PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT

Published May 6, 2020 · Recall initiated: March 11, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Merit Medical Systems, Inc.
Distributor
U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
Lot
SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
Risk classification
Class II
Area (FDA)
device
Country
United States