PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
Published May 6, 2020 · Recall initiated: March 11, 2020
Summary
FDA records medical device Class II recall Z-1796-2020, published May 6, 2020 and initiated March 11, 2020. Recalling firm: Merit Medical Systems, Inc.; Status: Terminated; Reported quantity: 1932; Lot: SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Merit Medical Systems, Inc.
- Distributor
- U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
- Lot
- SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAScout MD, REF MDS05A-01, MDS05C-01, MDS10C-01 Contents: (1) Scout MD Delivery System, Standard for use withOctober 21, 2026
- FDAEmbosphere 500-700 micrometer 2 mL Syringe Sterile Steam, REF: S420GH/EUA, S620GH/EUA; Embosphere 700-900October 7, 2026
- FDASplashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669June 24, 2026
- FDAPhilips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLYMay 6, 2026
- FDALunar iDXA Advance (iDXA with enCORE v16) bone densitometerMay 6, 2020
- FDAMoXy Laser Fiber, Material Number 0010-2400 - Product Usage: The fiber can be used for the surgicalMay 6, 2020
Questions about this recall
Was PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT recalled?
Yes. FDA published recall Z-1796-2020 on May 6, 2020. Reported hazard: During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
What should I do if I have this product?
Voluntary: Firm initiated · Letter