FDAFDA · Food and drugs·Recall number Z-1806-2013

ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or intervertebral body fusion to aid in the surgical correction or stabilization of the spine. The Aesculap PEEK Spinal Implant System is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.

Published July 31, 2013 · Recall initiated: June 18, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Aesculap Implant Systems, LLC initiated a recall of the Aesculap Prospace PEEK 5 degree Implant - 8 x 8.5 x 22 mm (SN038P) because the device is incorrectly etched with an 8 degree lordosis instead of a 5 degree lordosis. The device is labeled correctly, only the etching of the implant may be incorrect.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Aesculap, Inc.
Distributor
US Distribution including the states of NV and MI
Lot
Part no. SN038P, batch number 51915765
Risk classification
Class II
Area (FDA)
device
Country
United States