ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or intervertebral body fusion to aid in the surgical correction or stabilization of the spine. The Aesculap PEEK Spinal Implant System is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.
Published July 31, 2013 · Recall initiated: June 18, 2013
Summary
FDA records medical device Class II recall Z-1806-2013, published July 31, 2013 and initiated June 18, 2013. Recalling firm: Aesculap, Inc.; Status: Terminated; Reported quantity: 9; Lot: Part no. SN038P, batch number 51915765. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Aesculap Implant Systems, LLC initiated a recall of the Aesculap Prospace PEEK 5 degree Implant - 8 x 8.5 x 22 mm (SN038P) because the device is incorrectly etched with an 8 degree lordosis instead of a 5 degree lordosis. The device is labeled correctly, only the etching of the implant may be incorrect.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Aesculap, Inc.
- Distributor
- US Distribution including the states of NV and MI
- Lot
- Part no. SN038P, batch number 51915765
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAColumbus Revision Knee System, EnduRo Knee System Product Usage: The Columbus Revision Knee System and EnduRoMarch 9, 2016
- FDAFlexible Screw Driver SJ706R; Non Sterile; B Braun; Aesculap; Aesculap AG These instruments are intended forDecember 30, 2015
- FDAS4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with aAugust 12, 2015
- FDAMiethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinalFebruary 4, 2015
- FDABBL Campylobacter CSM Agar (Charcoal-Based Selective Medium), Catalog # 299614, packaged in cartons of 20July 31, 2013
- FDASwift Mobil shower chair, Swift Mobil 160 shower chair, and Swift Mobil Tilt shower chair. Height adjustableJuly 31, 2013
Questions about this recall
Was ProSpace Peek Implant 5 degree x 8.5 x 22 mm The device is intended for vertebral body replacement or intervertebral body fusion to aid in the surgical correction or stabilization of the spine. The Aesculap PEEK Spinal Implant System is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. recalled?
Yes. FDA published recall Z-1806-2013 on July 31, 2013. Reported hazard: Aesculap Implant Systems, LLC initiated a recall of the Aesculap Prospace PEEK 5 degree Implant - 8 x 8.5 x 22 mm (SN038P) because the device is incorrectly etched with an 8 degree lordosis instead of a 5 degree lordosis. The device is labeled correctly, only the etching of the implant may be incorrect.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail