Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: M003450200, Product is manufactured by Boston Scientific Corporation, Fremont, CA and distributed by Stryker Neurovascular, Fremont, CA The Neuroform Microdelivery Stent System is intended for use with embolic coils for the treatment of wide neck, intracranial, saccular aneurysms arising from a parent vessel with a diameter of greater than equal to 2 mm and greater than equal to 4.5 mm that are not amenable to treatment with surgical clipping.

Published July 31, 2013 · Recall initiated: June 24, 2013

Summary

FDA records medical device Class II recall Z-1808-2013, published July 31, 2013 and initiated June 24, 2013. Recalling firm: Stryker Neurovascular; Status: Terminated; Reported quantity: 23 units; Lot: Code: M003450200, Lot 15391019, Expiration Date: 2015-07. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same lot number. The Model Number on the outer box states UPN M003E345020 (4.5mm x 20 mm Neuroform 3 Stent) while the pouch label states UPN M003E3450300 (4.5mm x 30 mm Neuroform 3 Stent).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Neurovascular
Distributor
International Distribution including Austria, China, and Germany.
Lot
Code: M003450200, Lot 15391019, Expiration Date: 2015-07
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Stryker brand Neuroform 3 Microdelivery Stent System, 4.5 mm X 20 mm, REF Catalog Number: E345020, Model/Product Number: M003450200, Product is manufactured by Boston Scientific Corporation, Fremont, CA and distributed by Stryker Neurovascular, Fremont, CA The Neuroform Microdelivery Stent System is intended for use with embolic coils for the treatment of wide neck, intracranial, saccular aneurysms arising from a parent vessel with a diameter of greater than equal to 2 mm and greater than equal to 4.5 mm that are not amenable to treatment with surgical clipping. recalled?

Yes. FDA published recall Z-1808-2013 on July 31, 2013. Reported hazard: The device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same lot number. The Model Number on the outer box states UPN M003E345020 (4.5mm x 20 mm Neuroform 3 Stent) while the pouch label states UPN M003E3450300 (4.5mm x 30 mm Neuroform 3 Stent).

What should I do if I have this product?

Voluntary: Firm initiated · Letter