FDAFDA · Food and drugs·Recall number Z-1809-2012

T5 Hytrel Zipper Toga. Product Usage: The togas are intended to provide a barrier between the operating room and the members of the surgical team in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

Published June 20, 2012 · Recall initiated: June 1, 2012

Summary

FDA records medical device Class II recall Z-1809-2012, published June 20, 2012 and initiated June 1, 2012. Recalling firm: Stryker Instruments Div. of Stryker Corporation; Status: Terminated; Reported quantity: 23,653 ( past 3 years) 47937 all years; Lot: Part 0400-821-100, 0400-841-100 , 0400-851-100 0400-851-100S1, and 0400-851-100S2 all lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The clear tape, that is intended to aid in the prevention of patients fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process. Stryker has received 5 complaints associated with the defective seams which allowed bleed through of bodily fluids.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Instruments Div. of Stryker Corporation
Distributor
Worldwide Distribution - US (nationwide) including the states of: AK, AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the country of SWITZERLAND
Lot
Part 0400-821-100, 0400-841-100 , 0400-851-100 0400-851-100S1, and 0400-851-100S2 all lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was T5 Hytrel Zipper Toga. Product Usage: The togas are intended to provide a barrier between the operating room and the members of the surgical team in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms. recalled?

Yes. FDA published recall Z-1809-2012 on June 20, 2012. Reported hazard: The clear tape, that is intended to aid in the prevention of patients fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process. Stryker has received 5 complaints associated with the defective seams which allowed bleed through of bodily fluids.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit