Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional fluoroscopic x-ray system System code: 722079

Published May 6, 2020 · Recall initiated: April 3, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips North America, LLC
Distributor
Worldwide distribution. US states of FL, ID, IL, IN, MA, and OH, Canada, Australia, Austria, Belgium, Germany, and Netherlands New Zealand, South Africa, Switzerland, Thailand
Lot
Software:Philips Azurion 2.0 Interventional Fluoroscopic X-ray system with Interventional Workspot 1.5 US Serial Numbers: 637, 200, 909, 1276, 1309, 987 Canada: 1016 ROW: 11169 1129 1143 971 1135 824 934 1089 1008 1165 1080 1019 1248 523 1172 Expanded Recall OUS: 1223, 1330, 703424
Risk classification
Class II
Area (FDA)
device
Country
United States