Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.
Published May 6, 2020 · Recall initiated: March 18, 2020
Summary
FDA records medical device Class II recall Z-1810-2020, published May 6, 2020 and initiated March 18, 2020. Recalling firm: WELCH ALLYN, INC/MORTARA; Status: Terminated; Reported quantity: 78,521 units total; Lot: Model Numbers: ELI10-XXX-XXXXX, ELI350-XXX-XXXXX, ELI150C-XXX-XXXXX, ELI250C-XXX-XXXXX, ELI230-XXX-XXXXX, ELI280-XXX-XXXXX, ELI380-XYYZZ, RSCRIBE-xxx-xxxxx and CC-XXX-XXXX. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- WELCH ALLYN, INC/MORTARA
- Distributor
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
- Lot
- Model Numbers: ELI10-XXX-XXXXX, ELI350-XXX-XXXXX, ELI150C-XXX-XXXXX, ELI250C-XXX-XXXXX, ELI230-XXX-XXXXX, ELI280-XXX-XXXXX, ELI380-XYYZZ, RSCRIBE-xxx-xxxxx and CC-XXX-XXXX.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAX12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHAMay 6, 2020
- FDASurveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBLMay 6, 2020
- FDAReplaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for variousMay 6, 2020
- FDAPhilips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the AzurionMay 6, 2020
- FDAProACT Implantation Instrument Set, is a reusable, steam sterilizable surgical instrument set consisting of aMay 6, 2020
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is anMay 6, 2020
Questions about this recall
Was Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms. recalled?
Yes. FDA published recall Z-1810-2020 on May 6, 2020. Reported hazard: Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter