X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP, PAT CBL 4WIRE AHA SNAP HC.

Published May 6, 2020 · Recall initiated: March 18, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
WELCH ALLYN, INC/MORTARA
Distributor
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Lot
Model Numbers: X12PLUS-XXX-XXXXX, 9293-017-50, 9293-017-51, 9293-026-50, 9293-026-51, 9293-034-50
Risk classification
Class II