X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP, PAT CBL 4WIRE AHA SNAP HC.

Published May 6, 2020 · Recall initiated: March 18, 2020

Summary

FDA records medical device Class II recall Z-1814-2020, published May 6, 2020 and initiated March 18, 2020. Recalling firm: WELCH ALLYN, INC/MORTARA; Status: Terminated; Reported quantity: 78,521 units total; Lot: Model Numbers: X12PLUS-XXX-XXXXX, 9293-017-50, 9293-017-51, 9293-026-50, 9293-026-51, 9293-034-50. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
WELCH ALLYN, INC/MORTARA
Distributor
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Lot
Model Numbers: X12PLUS-XXX-XXXXX, 9293-017-50, 9293-017-51, 9293-026-50, 9293-026-51, 9293-034-50
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP, PAT CBL 4WIRE AHA SNAP HC. recalled?

Yes. FDA published recall Z-1814-2020 on May 6, 2020. Reported hazard: Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

What should I do if I have this product?

Voluntary: Firm initiated · Letter