Back to search
FDAFDA · Food and drugs·Recall number Z-1816-2020

Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable applications.

Published May 6, 2020 · Recall initiated: March 18, 2020

Summary

FDA records medical device Class II recall Z-1816-2020, published May 6, 2020 and initiated March 18, 2020. Recalling firm: WELCH ALLYN, INC/MORTARA; Status: Terminated; Reported quantity: 78,521 units total; Lot: Model Numbers: SUR12-XXX-XXXXX, SUR19-XXX-XXXXX, 9293-033-52, 9293-033-53. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
WELCH ALLYN, INC/MORTARA
Distributor
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Lot
Model Numbers: SUR12-XXX-XXXXX, SUR19-XXX-XXXXX, 9293-033-52, 9293-033-53.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable applications. recalled?

Yes. FDA published recall Z-1816-2020 on May 6, 2020. Reported hazard: Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

What should I do if I have this product?

Voluntary: Firm initiated · Letter