Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation.
Published August 7, 2013 · Recall initiated: March 28, 2013
Summary
FDA records medical device Class II recall Z-1819-2013, published August 7, 2013 and initiated March 28, 2013. Recalling firm: Custom Spine, Inc.; Status: Terminated; Reported quantity: 18 units; Lot: Catalog Number 048545; Lot Number L363. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Custom Spine, Inc.
- Distributor
- Nationwide Distribution including Nebraska and Virginia.
- Lot
- Catalog Number 048545; Lot Number L363
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAViSi Mobile Chest Sensor (3 lead-wire ECG, AAMI), Catalog No. 93-10014 The ViSi Mobile Monitoring System isAugust 7, 2013
- FDALPS Lower Extremity Dovetail Intercalary component. Intended for use in replacement of the mid-shaft portionAugust 7, 2013
- FDAViSi Mobile Monitor, Catalog No. 92-10010 The ViSi Mobile Monitoring System is intended for use by cliniciansAugust 7, 2013
- FDASynthes Synapse SD Drill Guide, Part Number SD389.477, Lot Number 6913900 Indicated for Hooks, Plate/Rods,August 7, 2013
- FDASynthes Large External Fixation, MR Conditional, Tube to Tube Clamp, Part Number 390.007, Lot Number 6800599August 7, 2013
- FDAMedline Steam Biological Indicators, SteriTec Product Code PL 353, Distributor Product Code MDS200525 EZTestAugust 7, 2013
Questions about this recall
Was Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation. recalled?
Yes. FDA published recall Z-1819-2013 on August 7, 2013. Reported hazard: On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail