Back to search
FDAFDA · Food and drugs·Recall number Z-1821-2012

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Published June 27, 2012 · Recall initiated: May 11, 2012

Summary

FDA records medical device Class II recall Z-1821-2012, published June 27, 2012 and initiated May 11, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 6; Lot: All systems with Eleva software version 4.x and stitching option. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Healthcare Inc.
Distributor
6 US customers
Lot
All systems with Eleva software version 4.x and stitching option.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system recalled?

Yes. FDA published recall Z-1821-2012 on June 27, 2012. Reported hazard: If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

What should I do if I have this product?

Voluntary: Firm initiated · Letter