Back to search
FDAFDA · Food and drugs·Recall number Z-1823-2012

Arrow International, Inc. 7FR 3-Lumen Arrowg+ard Blue Plus(R) Pressure Injectable Central Venous Catheter - Catalog number CDC-42703-XP1A The multiple-lumen pressure injectable catheter permits venous access to central circulation and injection of contrast media. The ARROWg+ard(R) antimicrobial surface is intended to provide protection against catheter-related infections.

Published June 27, 2012 · Recall initiated: June 7, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Arrow is recalling the product due to the possibility of voids in the catheter hub which could lead to an interlumen crossover.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Arrow International Inc
Distributor
USA (nationwide) AL, AZ, AR, CA, CO, FL, GA, IL, IN, KS, LA, MA, MI, MO, NV, NH, NC, OH, PA, TX, VA, WA, and WI.
Lot
Lot number RF2033361
Risk classification
Class II
Area (FDA)
device
Country
United States