Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US
Published May 22, 2024 · Recall initiated: April 2, 2024
Summary
FDA records medical device Class II recall Z-1823-2024, published May 22, 2024 and initiated April 2, 2024. Recalling firm: Fisher & Paykel Healthcare, Ltd.; Status: Ongoing; Reported quantity: 7,147; Lot: REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fisher & Paykel Healthcare, Ltd.
- Distributor
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
- Lot
- REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- New Zealand
Related recalls
- FDAPT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended toOctober 2, 2024
- FDANovalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours)May 22, 2024
- FDASyntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venousMay 22, 2024
- FDAFisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel; Healthcare CosyCot and KitNovember 19, 2014
- FDAInfant Nasal CPAP (continuous positive airway pressure) Prong, Nasal. CPAP Prongs (10 pack): Model No.May 28, 2014
- FDAFisher & Paykel Healthcare ICON CPAP (continuous positive airway pressure). Models: ICON Auto: ICONAAN,June 5, 2013
Questions about this recall
Was Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US recalled?
Yes. FDA published recall Z-1823-2024 on May 22, 2024. Reported hazard: Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter