Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

Published May 22, 2024 · Recall initiated: April 2, 2024

Summary

FDA records medical device Class II recall Z-1823-2024, published May 22, 2024 and initiated April 2, 2024. Recalling firm: Fisher & Paykel Healthcare, Ltd.; Status: Ongoing; Reported quantity: 7,147; Lot: REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fisher & Paykel Healthcare, Ltd.
Distributor
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Lot
REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017
Risk classification
Class II
Area (FDA)
device
Country
New Zealand

Related recalls

Questions about this recall

Was Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US recalled?

Yes. FDA published recall Z-1823-2024 on May 22, 2024. Reported hazard: Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.