FDAFDA · Food and drugs·Recall number Z-1830-2024
Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117
Published May 22, 2024 · Recall initiated: April 22, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom.
- Lot
- UDI-DI: (01)00884838099739(21) Serial Numbers: 42029 42217 42246 71044 71774 85146 85645 85663 85664 85668 85677 85680 85681 85685 85700 85704 85705 85706 85713
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States