Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain.
Published August 7, 2013 · Recall initiated: July 26, 2012
Summary
FDA records medical device Class II recall Z-1836-2013, published August 7, 2013 and initiated July 26, 2012. Recalling firm: St. Jude Medical; Status: Terminated; Reported quantity: 34,617; Lot: Multiple lots recalled are the ones with supplier's inner battery lots of 326-508. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- St. Jude Medical
- Distributor
- Worldwide Distribution--USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Belgium, Brazil, canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Saudi, South Africa, Spain, Sweden, Switzerland, Turkey and United Kingdom.
- Lot
- Multiple lots recalled are the ones with supplier's inner battery lots of 326-508
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACardioMEMS Heart Failure (HF) System that consist of: -CardioMEMS Hospital System Model CM3100 -CardioMEMSApril 2, 2025
- FDASt. Jude Medical Agilis NxT Steerable Introducer, REF 408309July 24, 2024
- FDACardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System providesMarch 15, 2023
- FDACardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System providesMarch 15, 2023
- FDAAU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model NumbersAugust 7, 2013
- FDAOmnicycle Elite; US Importer and Distributor: Accelerated Care Plus, Reno NV. Manufactured by MedicaAugust 7, 2013
Questions about this recall
Was Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain. recalled?
Yes. FDA published recall Z-1836-2013 on August 7, 2013. Reported hazard: As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne
What should I do if I have this product?
Voluntary: Firm initiated · Letter