Back to search
FDAFDA · Food and drugs·Recall number Z-1851-2014

RUSCH, LaserTube (Rubber), Laser Resistant Tracheal Tube, cuffed, Teleflex Medical, Willy-Rusch, GmbH. A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

Published July 2, 2014 · Recall initiated: May 23, 2014

Summary

FDA records medical device Class II recall Z-1851-2014, published July 2, 2014 and initiated May 23, 2014. Recalling firm: Teleflex Medical; Status: Terminated; Reported quantity: 1658 ea; Lot: Product Code: 102004040, Lot #13371, 13381, 13291, 13281, 13251 & 13231; Product Code: 102004050, Lot #13461, 13441, 13401, 13431, 13421, 13381, 13371, 1329, 13301, 13311, 13361, 13261, 13231, 13251 & 13241; Product Code: 102004060, Lot #13441, 13451, 13431, 13411, 13381, 13361, 13311, 13301, 13281, 13271, 13261, 13241 & 13231; Product Code: 102004070, Lot #13441, 13421, 13381, 13351, 13291, 13311, 13271 & 13261; Product Code: 102004070, Lot #13251 & 13241; and Product Code: 102004080, Lot #13371, 13301 & 13251. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Although the Natural Rubber Latex caution symbol is clearly displayed on the product label, the accompanying caution statement "Caution: This product contains Natural Rubber Latex which may cause Allergic Reactions" was not printed on the product label.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Teleflex Medical
Distributor
Nationwide Distribution and the country of Colombia.
Lot
Product Code: 102004040, Lot #13371, 13381, 13291, 13281, 13251 & 13231; Product Code: 102004050, Lot #13461, 13441, 13401, 13431, 13421, 13381, 13371, 1329, 13301, 13311, 13361, 13261, 13231, 13251 & 13241; Product Code: 102004060, Lot #13441, 13451, 13431, 13411, 13381, 13361, 13311, 13301, 13281, 13271, 13261, 13241 & 13231; Product Code: 102004070, Lot #13441, 13421, 13381, 13351, 13291, 13311, 13271 & 13261; Product Code: 102004070, Lot #13251 & 13241; and Product Code: 102004080, Lot #13371, 13301 & 13251.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was RUSCH, LaserTube (Rubber), Laser Resistant Tracheal Tube, cuffed, Teleflex Medical, Willy-Rusch, GmbH. A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway. recalled?

Yes. FDA published recall Z-1851-2014 on July 2, 2014. Reported hazard: Although the Natural Rubber Latex caution symbol is clearly displayed on the product label, the accompanying caution statement "Caution: This product contains Natural Rubber Latex which may cause Allergic Reactions" was not printed on the product label.

What should I do if I have this product?

Voluntary: Firm initiated · Letter