Back to search
FDAFDA · Food and drugs·Recall number Z-1858-2017

various polyethylene implants BIO PLUG 10MM Product Usage: Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant.

Published April 26, 2017 · Recall initiated: February 17, 2017

Summary

FDA records medical device Class II recall Z-1858-2017, published April 26, 2017 and initiated February 17, 2017. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 28,253 in total; Lot: Item 163005 Lot 309030. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer Biomet, Inc.
Distributor
Worldwide - US Nationwide in the ARGENTINA AUSTRALIA CANADA CHILE CHINA COSTA RICA ECUADOR GSCC INDIA ITALY JAPAN KOREA, THE REP OF (SOUTH) MALAYSIA MEXICO PANAMA Singapore SWITZERLAND THAILAND UK
Lot
Item 163005 Lot 309030
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was various polyethylene implants BIO PLUG 10MM Product Usage: Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant. recalled?

Yes. FDA published recall Z-1858-2017 on April 26, 2017. Reported hazard: Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.

What should I do if I have this product?

Voluntary: Firm initiated · Letter