various polyethylene implants BIO PLUG 10MM Product Usage: Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant.
Published April 26, 2017 · Recall initiated: February 17, 2017
Summary
FDA records medical device Class II recall Z-1858-2017, published April 26, 2017 and initiated February 17, 2017. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 28,253 in total; Lot: Item 163005 Lot 309030. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- Worldwide - US Nationwide in the ARGENTINA AUSTRALIA CANADA CHILE CHINA COSTA RICA ECUADOR GSCC INDIA ITALY JAPAN KOREA, THE REP OF (SOUTH) MALAYSIA MEXICO PANAMA Singapore SWITZERLAND THAILAND UK
- Lot
- Item 163005 Lot 309030
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: TheDecember 30, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. CupNovember 11, 2020
- FDAD-RAD, S.T. LOCKING SCREWS, T7, QTY: (1), STERILE R, in the following sizes: (a) 2.4MM X 10MM, RES 74692410April 26, 2017
- FDAvarious polyethylene implants Hips Active Articulation ArComXL Bearings Active Articulation E1 Bearings ArcosApril 26, 2017
Questions about this recall
Was various polyethylene implants BIO PLUG 10MM Product Usage: Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant. recalled?
Yes. FDA published recall Z-1858-2017 on April 26, 2017. Reported hazard: Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
What should I do if I have this product?
Voluntary: Firm initiated · Letter