Back to search
FDAFDA · Food and drugs·Recall number Z-1873-2026

AccuVu Angiographic Catheter, Straight Flush, Non-Braided, With 10 R/O Markers; ACCU-VU STR FL 4F X 100CM 035 NB 6SH 10 R/O 6@1CM 4@5CM; Catalog No.: 13709604; Product/UPN No.: H787137096045 (Box), H787137096040 (Pouch); Box Quantity: 5 pouches;

Published April 29, 2026 · Recall initiated: March 3, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Angiodynamics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, China, Chile, Curacao, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, New Zealand, Panama, Slovenia, South Africa, Spain, Sweden, Thailand.
Lot
ACCU-VU STR FL 4F X 100CM 035 NB 6SH 10 R/O 6@1CM 4@5CM; Catalog No.: 13709604; Product/UPN No.: H787137096045 (Box), H787137096040 (Pouch); UDI-DI: 25051684005996(Box), 15051684005999(Pouch); Lot No.: A2825044;
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card