FDAFDA · Food and drugs·Recall number Z-1875-2013
Mobiletta Mira Mobile X-ray System
Published August 14, 2013 · Recall initiated: May 2, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.
- Lot
- Model number 10273100
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States