ViperWire Advance with Flextip Peripheral Atherectomy Guide Wire, Part Number: 7-10026-01,Model Number: VPR-GW-FT14. Product Usage: The Stealth 360 Orbital PAD System and the Diamondback 360 Peripheral Orbital Atherectomy System are percutaneous orbital atherectomy systems indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
Published May 3, 2017 · Recall initiated: February 9, 2017
Summary
FDA records medical device Class II recall Z-1877-2017, published May 3, 2017 and initiated February 9, 2017. Recalling firm: Cardiovascular Systems Inc; Status: Terminated; Reported quantity: 595; Lot: Lot #: 175069, 175070, 176255. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
CSI is recalling three lots of VIPERWIRE ADVANCE WITH FLEXTIP PERIPHERAL ATHERECTOMY GUIDEWIRE . It was discovered that 0.018in ViperWire Advance with Flextip product was packaged as a 0.014in ViperWire Advance with Flextip.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cardiovascular Systems Inc
- Distributor
- US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, and WA
- Lot
- Lot #: 175069, 175070, 176255
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPeripheral Diamondback 1.50 Solid OAD, a percutaneous orbital atherectomy system indicated for use as therapyFebruary 14, 2018
- FDACSI, Orbital Atherectomy System (OAS) Saline Infusion Pump, Rx only, Part Number: 7-10014, Model Number:May 31, 2017
- FDA18L6 HD transducer on the ACUSON S Family ultrasound systems with software versions VD10A or VD10C; ModelMay 3, 2017
- FDAPressure Injectable Arrowg+ard Blue Plus Multi-Lumen Central Venous Catheterization Kit. Permits venous accessMay 3, 2017
- FDAHeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair theMay 3, 2017
- FDADIAMONDBACK 360 Peripheral 1.50 Classic, Model DBP-150CLASS145 Peripheral Orbital Atherectomy System (OAS) isMarch 1, 2017
Questions about this recall
Was ViperWire Advance with Flextip Peripheral Atherectomy Guide Wire, Part Number: 7-10026-01,Model Number: VPR-GW-FT14. Product Usage: The Stealth 360 Orbital PAD System and the Diamondback 360 Peripheral Orbital Atherectomy System are percutaneous orbital atherectomy systems indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. recalled?
Yes. FDA published recall Z-1877-2017 on May 3, 2017. Reported hazard: CSI is recalling three lots of VIPERWIRE ADVANCE WITH FLEXTIP PERIPHERAL ATHERECTOMY GUIDEWIRE . It was discovered that 0.018in ViperWire Advance with Flextip product was packaged as a 0.014in ViperWire Advance with Flextip.
What should I do if I have this product?
Voluntary: Firm initiated · Letter